eIFU Website Requirements: Setting Up a Site That Complies With EU 2021/2226
Short answer: an eIFU website that complies with Regulation (EU) 2021/2226 starts with a documented risk assessment under Article 4, then meets the conditions in Article 5 and the website rules in Article 7: a stable, directly accessible address, a format readable with free software, protection against tampering, minimal downtime, GDPR compliance, the right official languages, every issued version kept for 10 or 15 years, and free paper copies within 7 calendar days at the latest. Your QMS can hold the controlled IFU, but it does not meet these website obligations on its own.
The articles below are as amended by Regulation (EU) 2025/1234, in force since 16 July 2025; see what 2025/1234 changed.
Step 1: the eIFU risk assessment under Article 4
Every eIFU website starts with a documented risk assessment, because Article 4(1) requires one before instructions are supplied electronically instead of on paper. It must cover at least:
- the intended users' knowledge and experience, and their needs,
- the environment where the device is used,
- users' familiarity with the hardware and software needed to display the eIFU,
- access to the electronic resources needed at the time of use,
- safeguards protecting the data and content from tampering,
- safety and backup mechanisms if hardware or software fails,
- foreseeable medical emergencies needing information on paper,
- the impact of the website or internet being unavailable, and the measures to cope,
- the period within which paper copies are supplied on request,
- compatibility of the website with the devices used to display it,
- management of different IFU versions.
Under Article 4(2) the assessment is updated with post market experience, and under Article 5 point (1) it must show that electronic instructions maintain or improve the safety of paper ones.
eIFU website requirements in Article 5, point by point
Article 5 lists the conditions for supplying eIFU instead of paper. The ones that shape the website are:
- Point (2): provide the eIFU in all Member States where the device is made available or put into service, unless justified in the risk assessment.
- Point (3): a system to supply paper copies at no additional cost, within the period in the risk assessment and at the latest within 7 calendar days of a request, or with the delivery if requested at the time of order.
- Point (5): proper design and functioning of the eIFU, with verification and validation evidence.
- Point (7): software and hardware requirements for viewing, published in the catalogue or other device information.
- Point (8): a system that clearly shows when instructions were revised, and informs each user if a revision was needed for safety reasons.
- Points (9) and (10): retention, covered below.
- Point (11): instructions on the website in an official EU language determined by the Member State where the device is made available.
- Point (13): all issued electronic versions and their publication dates available on the website during the retention period, or obsolete versions on request.
Point (4) puts emergency information and, for built in displays, start up information on the device or a leaflet, outside the website.
eIFU website technical requirements in Article 7
Article 7(2) sets five requirements for any website holding instructions supplied electronically instead of on paper:
- a commonly used format readable with freely available software (point (a)),
- protection against unauthorised access and tampering of content (point (b)),
- server downtime and display errors reduced as far as possible (point (c)),
- compliance with the GDPR, Regulation (EU) 2016/679 (point (d)),
- a stable internet address, directly accessible for the whole retention period (point (e)).
Article 7(1) adds that where the eIFU ships on a storage medium or a built in display, it must also be on a website. Article 7(3) requires the address to be supplied to the EUDAMED UDI database once UDI registration applies. The label side is covered in the eIFU symbol.
eIFU retention periods
Retention depends on the device. Under Article 5 point (9), devices with a defined expiry date, other than implantables, keep their eIFU available for 10 years after the last device is placed on the market and at least 2 years after the expiry date of the last device produced. Under point (10), devices without an expiry date and implantable devices keep it for 15 years after the last device is placed on the market.
| Device | Keep eIFU available for | Source |
|---|---|---|
| Defined expiry date, not implantable | 10 years after last placed on the market, and at least 2 years after last expiry date | Article 5 point (9) |
| No defined expiry date | 15 years after last placed on the market | Article 5 point (10) |
| Implantable | 15 years after last placed on the market | Article 5 point (10) |
Validating the eIFU website
Validation is required by Article 5 point (5): you must ensure proper design and functioning and hold verification and validation evidence. Practical evidence covers the paper request workflow end to end, correct version and language per product, display on the devices named in the risk assessment, access controls on the publishing side, and availability monitoring.
Can a QMS host your eIFU website?
A QMS document module can be the source of truth for the IFU, but it does not on its own meet the eIFU website requirements. The regulation governs what users can reach, not where the master file lives. A QMS built for internal control normally sits behind a login, while Article 7(2)(e) requires an address that is directly accessible to users for 10 or 15 years. The public site also has to show the version list with publication dates (Article 5 point 13), serve the right language per Member State (point 11), take and fulfil paper requests within 7 calendar days (point 3), and keep downtime as low as possible (Article 7(2)(c)).
The common pattern is to approve the IFU in the QMS, then publish the released file to a public eIFU site that handles access, versions and paper requests. Dedicated eIFU services, Matrix eIFU among them, provide that public layer. (Matrix eIFU is made by the publisher of this guide.) See eIFU providers compared.
eIFU website FAQ
Do I need a separate eIFU platform or can my QMS host instructions for use?
Your QMS can be the source of truth for the approved IFU, but the public website obligations in Regulation (EU) 2021/2226 still apply: a stable, directly accessible address, every issued version with its publication date, the right languages, low downtime and paper requests within 7 calendar days. Most QMS document modules sit behind a login, so a public publishing layer is usually needed.
How long must an eIFU stay online?
Under Article 5 points (9) and (10) of Regulation (EU) 2021/2226: 10 years after the last device is placed on the market and at least 2 years after the last expiry date for devices with an expiry date, or 15 years for devices without one and for implantables.
How quickly must a paper IFU be supplied?
Within the period set in your Article 4 risk assessment and at the latest within 7 calendar days of the request, at no additional cost, under Article 5 point (3).
Does the eIFU website need to be validated?
Yes. Article 5 point (5) requires manufacturers to ensure the proper design and functioning of the eIFU and to provide verification and validation evidence.
Sources
- Regulation (EU) 2021/2226, consolidated text as of 16 July 2025
- Commission Implementing Regulation (EU) 2021/2226, original text
- Commission Implementing Regulation (EU) 2025/1234
- Regulation (EU) 2017/745 (MDR)
Related posts
- eIFU Symbol: Label Rules Under ISO 15223-1 and EU 2021/2226
The eIFU symbol explained: ISO 15223-1:2021 symbol 5.4.3, the eIFU indicator, and what Article 6 of Regulation (EU) 2021/2226 requires on the label. - eIFU Regulation 2025/1234: What Changed in EU 2021/2226
eIFU regulation update: what Regulation (EU) 2025/1234 changed in 2021/2226, article by article, with a before and after table. In force since 16 July 2025. - FDA Electronic Labeling for Devices: eIFU Rules Under 502(f)
FDA electronic labeling for medical devices: which devices may use eIFU under FD&C Act section 502(f), paper copies on request, home use, and 21 CFR 801.