Electronic Instructions for Use

FDA Electronic Labeling for Medical Devices: eIFU Rules for US Manufacturers

Short answer: section 502(f) of the Federal Food, Drug, and Cosmetic Act lets required labeling be made available solely by electronic means for prescription devices intended for use in health care facilities or by a health care professional, and for in vitro diagnostic devices intended for use by health care professionals or in blood establishments. The labeling must meet every other legal requirement, and the manufacturer must let users request it on paper and then provide it promptly at no additional cost. Devices outside those groups, such as those sold for lay use at home, are not covered by the provision.

The provision is short, and most of the work lies in the conditions around it. This page sets out the statutory text, its history, and what a US eIFU set up needs to do. For the EU rules, see eIFU website requirements.

What FDA electronic labeling law says

FDA electronic labeling rests on one sentence at the end of section 502(f) of the FD&C Act, codified at 21 U.S.C. 352(f). It allows required labeling for two groups to be made available solely by electronic means:

  • prescription devices intended for use in health care facilities or by a health care professional, and
  • in vitro diagnostic devices intended for use by health care professionals or in blood establishments.

Two conditions apply. The labeling must comply with all applicable requirements of law, and the manufacturer must afford users the opportunity to request the labeling in paper form and, after a request, promptly provide it without additional cost.

Where the FDA electronic labeling provision came from

The FDA electronic labeling sentence was added in 2002 and widened in 2004. Section 206 of the Medical Device User Fee and Modernization Act of 2002 (Public Law 107-250) added it, covering only prescription devices intended for use in health care facilities. The Medical Devices Technical Corrections Act (Public Law 108-214, 2004) extended it to prescription devices used by a health care professional and to professional and blood establishment IVDs, and changed the paper request right from "health care facilities" to "such users". The 2012 FDA Safety and Innovation Act amended other paragraphs of section 502 but not this sentence.

Which devices qualify for FDA electronic labeling

A device qualifies for FDA electronic labeling if it falls in one of the two statutory groups. For most devices that means it must be a prescription device. Under 21 CFR 801.109, a prescription device is one that is not safe except under the supervision of a practitioner licensed by law, and it is exempt from the "adequate directions for use" requirement of section 502(f)(1) only if its label bears "Rx only" (or the federal law caution statement) and its labeling carries the information practitioners need to use it safely.

DeviceElectronic only labeling under 502(f)?
Prescription device used in a hospital or clinicYes
Prescription device used by a health care professionalYes
IVD used by health care professionals or in blood establishmentsYes
Device sold over the counter for lay useNot covered
Prescription device used by the patient at homeNot covered by the "health care facilities or health care professional" wording

FDA electronic labeling and home use devices

Home use devices are not within the FDA electronic labeling sentence. Its wording is limited to use in health care facilities, use by a health care professional, and professional or blood establishment IVDs. A device whose intended user is a patient or carer at home does not fit any of those, so its required labeling cannot rely on 502(f) to be electronic only. Where a professional device also comes with patient labeling, plan on paper for the patient part.

Paper copies on request under FDA electronic labeling

A paper copy must be available to users who ask, and supplied promptly at no additional cost. The statute does not set a number of days. That is a contrast with the EU, where Article 5 point (3) of Regulation (EU) 2021/2226 sets an outer limit of 7 calendar days. A manufacturer selling in both markets can run one paper request process to the stricter EU limit and meet both.

Other 21 CFR 801 rules that still apply

Electronic delivery changes the medium, not the content rules in 21 CFR part 801. Two are easy to overlook on an eIFU site:

  • Revision date. 21 CFR 801.109(e) requires all labeling bearing information for use, other than labels and cartons, to bear the date of issuance or of latest revision. The online version must show it.
  • Symbols. Under 21 CFR 801.15(c)(1)(i), a symbol without adjacent English text must come from a standard (FDA recognised or another standards body's) used as specified, and be explained in a paper or electronic symbols glossary, with a prominent statement on or within the package saying where the glossary is. See the eIFU symbol.

What an FDA electronic labeling solution needs to do

A workable FDA electronic labeling solution for a US manufacturer does four things:

  1. Restrict scope by device. Publish electronic only labeling just for devices within 502(f), and keep paper in the box for everything else.
  2. Take and fulfil paper requests. A visible request route, a record of each request, and prompt free delivery.
  3. Control versions. Each online version is the same controlled labeling as the approved document, with its revision date under 801.109(e).
  4. Stay accessible. Users in a facility must be able to reach the current labeling when they need it.

Hosted eIFU services such as Matrix eIFU combine the website, version history and paper request handling. (Matrix eIFU is made by the publisher of this guide.) Compare options in eIFU providers compared.

FDA electronic labeling FAQ

Does the FDA allow electronic instructions for use?

Yes, for some devices. Section 502(f) of the FD&C Act allows required labeling to be made available solely by electronic means for prescription devices intended for use in health care facilities or by a health care professional, and for IVDs intended for use by health care professionals or in blood establishments.

Which law created FDA electronic labeling for devices?

Section 206 of the Medical Device User Fee and Modernization Act of 2002 added the electronic labeling sentence to section 502(f). The Medical Devices Technical Corrections Act of 2004 widened it to devices used by health care professionals and to professional IVDs.

How fast must a paper copy be sent in the US?

Section 502(f) says promptly and without additional cost, but sets no number of days. The EU limit under Regulation (EU) 2021/2226 is 7 calendar days at the latest.

Can home use devices have electronic only labeling in the US?

Section 502(f) does not cover them. Its electronic labeling sentence is limited to prescription devices used in health care facilities or by a health care professional, and to professional and blood establishment IVDs.

Sources

  1. 21 U.S.C. 352 (FD&C Act section 502) with amendment notes
  2. Public Law 107-250, Medical Device User Fee and Modernization Act of 2002
  3. Public Law 108-214, Medical Devices Technical Corrections Act (2004)
  4. 21 CFR Part 801, Labeling (eCFR)