eIFU Regulation (EU) 2025/1234: What Changed in 2021/2226, Article by Article
Short answer: Commission Implementing Regulation (EU) 2025/1234 amended the eIFU regulation, Regulation (EU) 2021/2226, and entered into force on 16 July 2025. It replaced the closed list of eligible devices in Article 3 with every MDR device intended for professional users, brought Annex XVI products into scope, deleted the duty to inform users who downloaded an IFU of every update (Article 5 point 12), let obsolete versions be supplied on request (Article 5 point 13), and added a duty to give the eIFU web address to the EUDAMED UDI database (Article 7(3)). It sets no separate transition period.
The amendment was adopted on 25 June 2025 and published in the Official Journal on 26 June 2025. Under its Article 2 it entered into force on the twentieth day after publication, which is 16 July 2025. Recitals (2) and (3) explain why: a Commission survey run from 1 August to 10 October 2024 found a clear preference among healthcare professionals for electronic instructions.
What 2025/1234 changed in the eIFU regulation: the table
The eIFU regulation changes are all in Article 1 of 2025/1234, which amends nine provisions of 2021/2226.
| Provision of 2021/2226 | Before | After 2025/1234 |
|---|---|---|
| Article 1, third subparagraph | Annex XVI products excluded | Deleted, so Annex XVI products are in scope |
| Article 2(2) and 2(3) | "Professional users" of the medical device; "fixed installed medical devices" and their accessories | "Medical" deleted from 2(2); 2(3) redefined as "fixed installed devices" |
| Article 3(1) | Only implantable and active implantable devices, fixed installed devices, and devices with a built in IFU display | Any device under Article 1(4) MDR intended for use by professional users |
| Article 3(2) | Exclusive professional use, and use by others not reasonably foreseeable | Where lay use is reasonably foreseeable, the lay person IFU must be on paper |
| Article 5 point (6) and (11) | "Medical devices"; language for "the user or patient" | Wording narrowed: "medical" and "to the user or patient" deleted |
| Article 5 point (12) | Systems to inform users who downloaded the IFU of updates or corrective actions | Deleted |
| Article 5 point (13) | All historical versions on the website | All issued versions and publication dates on the website for the retention period, or obsolete versions on request |
| Article 6(4) | Patient part of IFU for implantables not electronic | Deleted |
| Article 7(2)(f) | Previous versions and dates on the website | Deleted (now covered by Article 5 point 13) |
| Article 7(3) | Did not exist | Internet address supplied to the EUDAMED UDI database |
| Article 8 | Notified body review of Articles 4 to 7 | Deleted |
| Article 9, second paragraph | Website rules for eIFU supplied in addition to paper | Deleted |
eIFU regulation scope after 2025/1234
The eIFU regulation now covers every MDR device intended for professional users, instead of three named categories. Article 3(1), as replaced, allows electronic instructions instead of paper for devices referred to in Article 1(4) of Regulation (EU) 2017/745 intended for use by professional users. Recital (3) confirms this includes devices under the Article 120 transitional provisions, and recital (4) and the deletion of the third subparagraph of Article 1 bring in Annex XVI products intended for professional use.
The lay user limit stayed, in a new form. Article 3(2) now says that where it is reasonably foreseeable that a professional device is also used by lay persons, the instructions intended for lay persons must be provided on paper. Article 3(3), on software providing its IFU through the software itself, was not amended.
Article 5 changes in the eIFU regulation
Two Article 5 conditions changed in substance. Point (12), the duty to have systems to inform users who had downloaded instructions of updates or corrective actions, was deleted. Point (13) was replaced: during the retention periods in points (9) and (10), all issued electronic versions and their date of publication must be available on the website or, for obsolete versions, be made available upon request.
Everything else in Article 5 stands, including the documented risk assessment demonstrating equal or better safety (point 1), availability in every Member State where the device is sold unless justified (point 2), and the free paper copy within the risk assessment period and at the latest within 7 calendar days (point 3). Point (8), the duty to show when instructions were revised and to inform each user if the revision was needed for safety reasons, was not amended either.
EUDAMED and the eIFU regulation: Article 7(3)
Article 7(3) makes the eIFU web address registered data. At the latest from the date registration of devices in the UDI database applies under Article 123(3), point (d) or (e), of Regulation (EU) 2017/745, the manufacturer must supply the internet address from Article 7(2)(e) to the UDI database under Annex VI, Part B, point 22 of the MDR. That MDR field, "URL for additional information, such as electronic instructions for use", was optional before; for eIFU it now has to be filled.
Deletions that remove overlaps
Recital (7) explains the deletions as removing uncertainty and overlap. Article 8, which had the notified body review Articles 4 to 7, was deleted because compliance of the information supplied by the manufacturer is already part of conformity assessment under the MDR. Article 6(4) and the second paragraph of Article 9 were also deleted.
Transition dates in the 2025 eIFU regulation
Regulation (EU) 2025/1234 contains no transition period of its own. Article 2 sets only the entry into force date, and the act is binding and directly applicable in all Member States from 16 July 2025. The only deferred obligation is Article 7(3), which is tied to the date EUDAMED UDI registration becomes mandatory rather than to a fixed date.
For what these changes mean for your label and website, see the eIFU symbol and eIFU website requirements. Software options are covered in eIFU providers compared.
Regulation 2025/1234 FAQ
When did Regulation (EU) 2025/1234 enter into force?
On 16 July 2025, the twentieth day after its publication in the Official Journal on 26 June 2025, under its Article 2. It has no separate transition period.
Which devices can use eIFU after 2025/1234?
Devices under Article 1(4) of Regulation (EU) 2017/745 intended for professional users, including devices under the Article 120 transitional provisions and Annex XVI products for professional use. Instructions intended for lay persons must stay on paper where lay use is reasonably foreseeable.
Do I still have to notify users of IFU updates?
Article 5 point (12), the duty to inform users who downloaded the instructions of updates or corrective actions, was deleted. Article 5 point (8) still requires a system to show when instructions were revised and to inform each user if the revision was necessary for safety reasons.
Does 2025/1234 change the paper copy rule?
No. Article 5 point (3) still requires a free paper copy within the period in your risk assessment and at the latest within 7 calendar days of a request.
Sources
- Commission Implementing Regulation (EU) 2021/2226, original text
- Commission Implementing Regulation (EU) 2025/1234
- Regulation (EU) 2021/2226, consolidated text as of 16 July 2025
- Regulation (EU) 2017/745 (MDR)
Related posts
- eIFU Symbol: Label Rules Under ISO 15223-1 and EU 2021/2226
The eIFU symbol explained: ISO 15223-1:2021 symbol 5.4.3, the eIFU indicator, and what Article 6 of Regulation (EU) 2021/2226 requires on the label. - FDA Electronic Labeling for Devices: eIFU Rules Under 502(f)
FDA electronic labeling for medical devices: which devices may use eIFU under FD&C Act section 502(f), paper copies on request, home use, and 21 CFR 801. - eIFU Website Requirements: How to Comply With EU 2021/2226
eIFU website requirements under Regulation (EU) 2021/2226: risk assessment, Article 5 and 7 rules, languages, retention, 7 day paper requests, QMS hosting.